Standard configuration: ECG, blood oxygen, blood pressure, respiratory rate, temperature, heart rate.
Optional configuration: capacitive screen, printer, WIFI, mainstream/bypass ETCO2, Masimo blood oxygen, Suntech blood pressure, etc.
Accessories: five-lead heart wire * 1, blood oxygen probe * 1, blood pressure extension tube * 1, blood pressure cuff * 4, disposable electrode sheet * 15, power cord * 1.
Optional accessories: dollies, wall brackets, etc.
Linux operating system, based on multi-user, multi-tasking, multi-CPU operating system; good interactive interface as well as safe and stable operation flow; HD10VET animal monitor equipped with linux operating system to achieve multi-domain technology expansion; WIFI connection to the central monitoring station, as well as USB upgrade, expansion of the TV and other larger screen monitoring (optional), etc.
Display method seven-lead the same screen.
Heart rate measurement range: 20-350bpm.
Alarm range: 20-350bpm.
Common mode rejection ratio ≥105db.
Waveform speed 6.25, 12.5, 25, 50mm/s four levels of selection.
Gain selection Seven selection modes.
QRS volume with 5 levels of adjustment.
Monitoring mode Diagnostic, monitoring, surgery, ST four monitoring modes.
Heart rate arrhythmia analysis With 23 kinds of heart rate arrhythmia analysis.
Electrodebrider protection The ECG module can work normally in an electrosurgical interference environment with electrodebrider cutting power of 300W and coagulation power of 100W, and the accuracy is not affected.
Measurement range: 0 – 150 breaths/min.
Respiratory excitation waveform Sine wave signal, 62.8kHz (±10%), <500μA.
Asphyxiation alarm with more than 9 levels of alarm time limit setting.
Gain adjustment with 8 levels of gain adjustment.
Measurement range Blood pressure (units) Equine Canine Feline
Systolic blood pressure mmHg 40-255 40-200 40-135
kPa 5.3-34.0 5.3-26.7 5.3-18.0
Diastolic blood pressure mmHg 20-205 20-150 20-100
kPa 2.7-27.3 2.7-20.0 2.7-13.3
Mean pressure mmHg 27-220 27-165 27-110
kPa 3.6-29.3 3.6-22.0 3.6-14.7
Cuff pressure measurement range: 0 ~ 290±3mmHg.
Software overpressure protection Equine: no more than 293 mmHg
Canidae: no more than 250 mmHg.
Feline: not to exceed 148 mmHg.
Hardware overpressure protection Equidae/canines: up to 300 mmHg.
Feline: up to 150 mmHg.
Measurement interval in automatic mode 2.5 min, 5 min, 10 min, 15 min, 20 min, 30 min, 45 min, 60 min, 90 min, 120 min.
Measurement range: 69%-100%.
Pulse Tone With pulse tone switch function.
Waveform speed 3 or more waveform speed options.
Sensitivity 3 sensitivity options available.
HD10Vet animal monitor adopts dual body temperature module configuration, with dual-channel temperature difference display.
real-time display of the absolute value of the measured temperature difference.
Measurement range: 0mmHg-114mmHg
Compensation function with O2, automatic compensation for atmospheric pressure, N2O compensation
Measurement function: FiCO2, EtCO2, awRR